概述

  • 产品名称
    大鼠MCP1 ELISA试剂盒(CCL2)
    参阅全部 MCP1 试剂盒
  • 检测方法
    Colorimetric
  • 样品类型
    Cell culture extracts, Tissue Extracts
  • 检测类型
    Sandwich (quantitative)
  • 灵敏度
    < 15 pg/ml
  • 范围
    24.67 pg/ml - 18000 pg/ml
  • 复检

    93 %

    特定样本复检结果
    样品类型 平均% 范围
    Cell Lysate 93.73 84% - 105%
    Tissue Lysate 92.75 83% - 103%

  • 实验步骤
    Multiple steps standard assay
  • 种属反应性
    与反应: Rat
  • 产品概述

    Abcam’s MCP1 Rat ELISA (Enzyme-Linked Immunosorbent Assay) kit is an in vitro enzyme-linked immunosorbent assay for the quantitative measurement of Rat MCP1 in cell lysates and tissue lysates.

    This assay employs an antibody specific for rat MCP1 coated on a 96-well plate. Standards and samples are pipette into the wells and MCP1 present in a sample is bound to the wells by the immobilized antibody. The wells are washed and biotinylated anti-rat MCP1 antibody is added. After washing away unbound biotinylated antibody, HRP conjugated streptavidin is pipetted to the wells. The wells are again washed, a TMB substrate solution is added to the wells and color develops in proportion to the amount of MCP1 bound. The Stop Solution changes the color from blue to yellow, and the intensity of the color is measured at 450 nm.

  • 经测试应用
    适用于: Sandwich ELISAmore details
  • 平台
    Microplate

性能

应用

Our Abpromise guarantee covers the use of ab100778 in the following tested applications.

The application notes include recommended starting dilutions; optimal dilutions/concentrations should be determined by the end user.

应用 Ab评论 说明
Sandwich ELISA Use at an assay dependent concentration.

图片

  • Representative standard curve using ab100778

实验方案

文献

This product has been referenced in:
  • Murphy CJ  et al. Age- and species-dependent infiltration of macrophages into the testis of rats and mice exposed to mono-(2-Ethylhexyl) phthalate (MEHP). Biol Reprod 91:18 (2014). Sandwich ELISA ; Rat . Read more (PubMed: 24876407) »

See 1 Publication for this product

客户评价及客户问答

Thank you for your enquiry. I apologize for the delay in replying.

The two kits use the same antibodies but a few other reagents in each kit differ, to optimize the assay for different sample types. Product ab100777 is best for plasma, serum...

Read More

Thank you for contacting us. The kit requires that you use 100 ul of sample in each well. Please let me know if you have any further questions.

Please note: All products are "FOR RESEARCH USE ONLY AND ARE NOT INTENDED FOR DIAGNOSTIC OR THERAPEUTIC USE"

注册