Difference between normal group and lung adenocarcinoma (ADC) group was very clear.
But the absolute values measured by ab99992 were remarkably higher than clinical tests in hospital or CRO companies.
To make matters worse, when we increased the dilution factor (x2, x5, or x10 dilution), the absolute values of CEA also gradually increased.
Hence this kit can be used for only "relative quantification" studies.
But the absolute values measured by ab99992 were remarkably higher than clinical tests in hospital or CRO companies.
To make matters worse, when we increased the dilution factor (x2, x5, or x10 dilution), the absolute values of CEA also gradually increased.
Hence this kit can be used for only "relative quantification" studies.